| Circulation Method | Forced air circulation |
| Operating Temp. Range | Room temp. +10°C to 210°C |
| Temp. Adjustment Accuracy | ±1°C (at 210°C) |
| Temp. Distribution Accuracy | ±2.5°C (at 210°C) |
| Max. Temp. Reaching Time | ~60 min. |
| Interior/Exterior Material | Stainless steel / Cold rolled steel plate with melamine resin baking finish |
| Heat Insulating Material | Glass wool |
| Heater | SUS pipe heater/1.5kW x 2 |
| Fan Type / Fan Motor | 1pc / 30W |
| Cable Hole | 30mm I.D. (on the right side) 1pc. |
| Exhaust Port | 30mm I.D.x2 (the back) |
| Observation Window | None |
| Temp. Controller | 3 patterns program controller, PID control by microprocessor |
| Temp. Setting Method | Digital setting by UP/DOWN key |
| Temp. Display | Measurement temp. : Digital display by green LED / Setting temp. : Digital display by red LED |
| Timer | 1 min. to 99 Hrs. 59 min. and 100 Hrs. to 999 Hrs. 50 min. with timer wait function |
| Operation Function | Fixed temperature operation, Program operation, Auto start, Auto stop, Quick Auto-stop |
| Program Mode | Program operation : 3 patterns, 30 steps (30 stepsx1, 15 stepsx2, 10 stepsx3) Pattern repeat function |
| Additional Functions | Calibration off-set function, Key lock, Uninterruptible power for memory |
| Heater Circuit Control | SSR control |
| Sensor | K-thermocouple |
| Safety Device | Self diagnostic functions (temp. sensor abnormal, Heater disconnection, SSR- short, Automatic overheating prevention), Key lock function, Overheat prevention, Electric leakage breaker with over current protection |
| Internal Dimensions WxDxH | 600x500x1000 mm |
| External Dimensions WxDxH | 710x651x1608 mm |
| Internal Capacity | 300L |
| Shelf Plate With Standard Load | ~15kg/piece |
| Shelf Rest Step Number / Shelf Rest Pitch | 29 steps / 30mm |
| Power Source 50/60Hz | 220V 15A (no plug, round terminal) |
| Weight | ~110kg |
| Shelf Plate | Stainless steel, 1pc on the bottom screwed |
| Shelf Plate / Bracket | 4 pcs. / 8 pcs. |